BIOSTELLAR CO₂ Incubator for Cell and Tissue Culture
RYX Upgraded Series
The BIOSTELLAR CO₂ incubator provides a temperature-controlled, humidified environment for in vitro cell and tissue culture. Available in 50 L, 150 L and 240 L capacities, it combines infrared CO₂ sensing, six-sided heating and a 7-inch touchscreen. Separate chamber sterilization versions are available with 90°C moist heat, 140°C dry heat or 180°C dry heat.
Quantity
Features:
- Defined incubation conditions: Temperature adjustment from ambient +5°C to 60°C, with 0.1°C resolution, ±0.1°C fluctuation and temperature uniformity ≤0.3°C at 37°C.
- Infrared CO₂ measurement: An infrared sensor and PID control regulate CO₂ over a 0–20% range, with a specified concentration fluctuation of ±0.5%.
- Six-sided heating and gentle circulation: An air-jacket heating design and micro-airflow circulation distribute heat throughout the chamber. The fan stops automatically when the door opens.
- Water-pan humidification: A large-area water pan provides relative humidity ≥90% at 37°C through evaporation.
- Accessible chamber construction: A mirror-finish stainless steel interior with rounded corners supports cleaning, while an inner glass door allows observation without fully exposing the chamber.
- Configuration-specific sterilization: Choose a version with 90°C moist-heat, 140°C dry-heat or 180°C dry-heat chamber treatment.
- Touchscreen operation and data review: A 7-inch colour touchscreen supports fixed-value and programmed operation, scheduled start, real-time and historical curves, and USB export of historical data.
- Monitoring and protection: Temperature, CO₂ concentration and door alarms are combined with screen locking and power-failure protection.
Request a quote now
BS-WJ-3/BS-WJ-3-160 CO2 Incubator
BS-WJ-3/BS-WJ-3-160
Request a quote now
BS-WJ-2/BS-WJ-2-160 CO2 Incubator
BS-WJ-2/BS-WJ-2-160
Request a quote now
BS-LCI Series water jacket CO2 incubator 85L–270L
BS-LCI-85 / BS-LCI-165 / BS-LCI-270 Series
Request a quote now
BS-WJ-2 CO2 Incubator
BS-WJ-2