What Do “d” and “e” Mean on a Laboratory Balance?
Understanding “d” and “e” on a Laboratory Balance
When using a laboratory balance, some users may notice two values in the specifications or on the nameplate: d and e.
A common question is:
When the balance shows a weighing result, is it showing according to “d” or “e”?
The simple answer is:
The normal weighing display is based on “d”. “e” is mainly used for metrological verification.
They are two different technical parameters, not two weighing modes.
What Is “d”?
d = Actual Scale Interval / Display Division
It represents the smallest increment normally displayed by the balance.
For example, if:
d = 10 mg = 0.01 g
the balance can display:
100.00 g → 100.01 g → 100.02 g
So when you use the balance for normal weighing, the value shown on the screen is displayed according to d.
What Is “e”?
e = Verification Scale Interval
It is mainly used for metrological verification and evaluation of permissible weighing error.
For example, the balance shown above is marked:
Max: 4200 g
Min: 0.5 g
d = 10 mg (0.01 g)
e = 100 mg (0.1 g)
For this balance:
e = 10d
However, this does not mean that the balance has two weighing resolutions that the operator needs to select.
The balance still displays the normal weighing result according to d = 0.01 g, while e = 0.1 g is used for verification purposes.